COMPARISON OF ANALYTIC PERFORMANCE OF RESULTS OF DETERMINATION OF Pb METAL IONS USING ATOMIC ABSORPTION SPECTROPHOTOMETER of VARIOUS in SAMARINDA LABORATORIES

  • Arifin Syam Surya Arifin
  • Bohari Yusuf Jurusan Kimia FMIPA Universitas Mulawarman
  • Aman Sentosa Panggabean Jurusan Kimia FMIPA Universitas Mulawarman

Abstract

The study about the comparison of analytical performance of the determination of Pb metal ions using Atomic Absorption Spectrophotometer in various laboratories in samarinda has been conducted. The validated methods are linearity, detection limits, quantity and precision limits. The results showed, a standardized regression test obtained by laboratory A, a value of R2 ≥ 0.4103, laboratory B a value of R2 ≥ 0.983 and laboratory C, a value of R2 ≥ 0.000. The detection limit found at laboratory A was 0.6728 µg/L, laboratory B was 0.1989 µg/L and laboratory C was 0.0001 µg/L. The quantization limit obtained by laboratory A was 0.6817 µg/L, laboratory B was 0.6063 µg/L and laboratory C was 0.0001 µg/L. The precision test of laboratory A was 7.2208, laboratory B was 5.4557 and laboratory C was 0,000. The result of determining Pb of gas station soil samples in laboratory A is 0.737 µg/L and laboratory B is 0.391 µg/L, so that means they have exceeded the quality standards set by the Indonesian National Standard (INS).

Keyword
Method validation, plumbum, gas station soil, Atomic Absorption Spectrophotometer

Downloads

Download data is not yet available.

References

[1] Mustafa, M. (2012). Modul Pembelajaran Dasar-dasar Ilmu Tanah. Universitas Hasanuddin Makasar.
[2] Koestore. (1995). Kimia dan ekotoksilogi pencemaran. Jakarta: UI-Press
[3] Palar, H. 1994. Pencemaran dan Toksislogi Logam Berat. Penerbit PT. Rineka Cipta. Jakarta: Cetakan pertama.
[4] Alloway, B. (1995). Heavy Metals in Soils Blackie Academic & Professional. London.
[5] Panggabean A S, Silaban H S,Pasaribu S P, and Alimuddin (2019). ethod validation of Cd (II) determination in lubrication oilby direct dilution method using atomic absorption spectrophotometer. Samarinda: Journal of Physics
[6] Day, R.A dan Underwood, A. (1986). Analisis kimia kiantitatif (Edisi Kelima). Jakarta: Erlangga.
[7] Napitupulu R M, Julia R R D, Panggabean A S. (2019). Validasi metode penentuan mn dalam oli Lubrikan Dengan Metode Pengenceran Langsung Menggunakan Spektrofotometer Serapan Atom. Samarinda: Indonesian journal of Chemical Research Vol 6 No 2
[8] Suwansa S., Buchari B., Panggabean A. S. (2008). Pengembangan Metode Prakonsentrasi dengan Teknik Injeksi Air untuk Analisis Cu2+ dan Pb2+ dalam Air Aliran Sungai Citarum dan Waduk Saguling. Samarinda: Jurnal Matematika dan Sains Vil 13 No 3.
[9] Panggabean, A. S., Pasaribu, S., Bohari, Nurhasanah, 2014, Preconcentration of Chromium (VI) at Trace Levels Using Acid Alumina Resin With Column Method, Indones. J. Chem., 14(1), 51–56.
[10] Iskandar B., Panggabean A. S, Kartika R. (2017). Validasi Metode Penetuan Arsenik pada Sampel Air Sumur Bor dengan Menggunakan Spektrofotometer Serapan Atom di PT. Geoservices Balikpapan. Samarinda: Prosiding Seminar Nasional Kimia
[11] Badan Standarisasi Nasional. Standar Nasional Indonesia Bahan Beracun dan Berbahaya (B3), 2004.
Published
2021-11-30
How to Cite
SURYA, Arifin Syam; YUSUF, Bohari; PANGGABEAN, Aman Sentosa. COMPARISON OF ANALYTIC PERFORMANCE OF RESULTS OF DETERMINATION OF Pb METAL IONS USING ATOMIC ABSORPTION SPECTROPHOTOMETER of VARIOUS in SAMARINDA LABORATORIES. JURNAL KIMIA MULAWARMAN, [S.l.], v. 19, n. 1, p. 17-23, nov. 2021. ISSN 2476-9258. Available at: <http://jurnal.kimia.fmipa.unmul.ac.id/index.php/JKM/article/view/909>. Date accessed: 26 apr. 2024. doi: https://doi.org/10.30872/jkm.v19i1.909.