Abstract

Research on verification of the method for determining Cu and Zn metal ions using ICP-OES for seawater samples, namely AL-1 and AL-2, has been carried out. The verification value for Cu metal is Linearity = 1,000; Accuracy (% trueness) = 100.7; % bias = -0.74; Precision (% RSD = 1.47 ≤ 2/3 % CV Horwitz = 16.82); % Recovery = 100.2%; MDL = 0.01; LOQ = 0.05; Uncertainty value = 0.03; Low repeatability (% RSD = 2.17 < 0.5 CV Horwitz = 7.13); middle (% RSD = 0.67 < 0.5 CV Horwitz = 6.72); high (% RSD = 0.72 < 0.5 CV Horwitz = 5.85); Reproducibility (horrat value) low = 0.58; middle = 0.38 and high = 0.34. The verification value of Zn metal is Linearity = 1,000; Accuracy (% trueness) = 95.1; % bias = 4.93; Precision (% RSD = 10.10 ≤ 2/3 % CV Horwitz = 16.94); % Recovery = 95.6%; MDL = 0.01; LOQ = 0.05; Uncertainty value = 0.04. Low repeatability (% RSD = 1.17 < 0.5 CV Horwitz = 7.17); middle (% RSD = 0.69 < 0.5 CV Horwitz = 6.28); high (% RSD = 0.70 < 0.5 CV Horwitz = 5.85). Reproducibility (horrat value) low concentration= 0.57; medium concentration = 0.50; high concentration = 0.33. The sample concentrations and uncertainties of Cu and Zn metals are AL-1 (0.25 ± 0.03) and (0.01 ± 0.04); AL-2 (0.30 ± 0.03) and (0.40 ± 0.04). The measurement results show that the verification parameter values are requirements for determining metals in seawater.

Downloads

Download data is not yet available.

References

[1] D. R. Pratama, M. Yusuf, and M. Helmi, “Kajian kondisi dan sebaran kualitas air di perairan selatan Kabupaten Sampang, Ja-Tim,” Jurnal Oseanografi, vol. 5, no. 4, pp. 479–488, 2016. [Online]. Available: https://ejournal.pnc.ac.id/index.php/jppl/article/view/2161/729.

[2] Pemerintah Republik Indonesia, “Peraturan Pemerintah (PP) Nomor 22 Tahun 2021 tentang Penyelenggaraan Perlindungan dan Pengelolaan Lingkungan Hidup,” 2021. [Online]. Available: https://peraturan.bpk.go.id

[3] Mukamto, E. N. Hidayah, and Susilowati, “Kajian Kualitas Air Laut dan Sebaran Indeks Pencemaran di Pesisir Utara Kecamatan Palang-Tuban pada Musim Kemarau,” Jurnal Pengendalian Pencemaran Lingkungan (JPPL), vol. 6, no. 1, Mar. 2024, doi: 10.35970/jppl.v6i1.2161.

[4] N. Khairunnisa, I. Mandang, and R. Munir, “Penentuan status mutu air laut menggunakan metode indeks pencemaran di perairan Bontang Kalimantan Timur,” Jurnal Geosains Kutai Basin, vol. 7, no. 1, pp. 1–12, 2024, doi: 10.30872/geofisunmul.v7i1.955.

[5] A. S. Panggabean, R. Malinda, S. Pasaribu, and T. Wirawan, “Verification of the test method for determination of boron in NPK blending fertilizer using inductively coupled plasma-optical emission spectrometry (ICP-OES),” AIP Conf. Proc., vol. 2360, 2021, doi: 10.1063/5.0059480.

[6] A. Hadi, Verifikasi Metode Pengujian Air dan Air Limbah: Mendukung Penerapan ISO/IEC 17025:2017. Bogor, Indonesia: IPB Press, 2020.

[7] P. Pirdaus, M. Rahman, Rinawati, N. L. G. R. Juliasih, D. Pratama, and A. A. Kiswandono, “Verifikasi metode analisis logam Pb, Cd, Cr, Cu, Ni, Co, Fe, Mn dan Ba pada air menggunakan inductively coupled plasma-optical emission spectrometer (ICP-OES),” Analit: Analytical and Environmental Chemistry, vol. 3, no. 1, pp. 1–10, Apr. 2018, doi: 10.23960/aec.v3.i1.2018.p1-10.

[8] A. Susanto, T. Mulyani, and S. Nugraha, “Validasi metode analisis penentuan kadar logam berat Pb, Cd dan Cr terlarut dalam limbah cair industri tekstil dengan metode inductively coupled plasma optical emission spectrometry Prodigy7,” Jurnal Ilmu Lingkungan, vol. 19, no. 1, pp. 191–200, 2021, doi: 10.14710/jil.19.1.191-200.

[9] Sujito, S. P. Pasaribu, T. Wirawan, and A. S. Panggabean, “Gas-liquid separator modified on CV-AAS system for determination of total Hg in the industrial wastewater samples,” Molekul, vol. 17, no. 3, pp. 292–300, 2022, doi: 10.20884/1.jm.2022.17.3.5039.

[10] L. H. Simanjuntak, E. Erwin, and A. S. Panggabean, “Prakonsentrasi ion Cd(II) dalam sampel pupuk nitrogen, fosfor dan kalium (NPK) menggunakan alumina teraktivasi sebagai material pengisi kolom,” ALCHEMY Jurnal Penelitian Kimia, vol. 16, no. 2, pp. 1–11, 2020, doi: 10.20961/alchemy.16.2.39363.1-11.

[11] E. Damastuti, S. Kurniawati, and N. Adventini, “Komparasi metode analisis aktivasi neutron dengan spektrometri serapan atom pada analisis unsur atom Zn dalam sampel makanan,” Sains dan Teknologi Nuklir, pp. 326–333, 2009.

[12] R. Anggraini, R. Hairani, and A. S. Panggabean, “Validasi metode penentuan Hg pada sampel waste water treatment plant dengan menggunakan teknik bejana uap dingin-spektrofotometer serapan atom (CV-AAS),” Jurnal Kimia Mulawarman, vol. 16, no. 1, pp. 10–15, Nov. 2018.

[13] Riyanto, Validasi dan Verifikasi Metode Uji Sesuai dengan ISO/IEC 17025 Laboratorium Pengujian dan Kalibrasi. Yogyakarta, Indonesia: Deepublish, 2014.

[14] A. R. Utami, “Verifikasi metode pengujian sulfat dalam air dan air limbah sesuai SNI 6989.20:2009,” Teknologi Proses dan Inovasi Industri, vol. 2, no. 1, pp. 19–25, 2017.

[15] A. Hadi, Ketidakpastian Pengujian: Mendukung Penerapan ISO/IEC 17025:2017. Bogor, Indonesia: IPB Press, 2019.